|
|||||||||||||||
|
|
Cartridge Dust Collector Historyby Gary Berwick, P. Eng.
Excerpts from an article published in various
media. |
||||||||||||||
![]() |
Referring to the Figure 1:
|
... forces developed during the pulse cleaning process gave the gaskets a permanent set which shortened the filter element life dramatically.
... Many times the cleaning forces would break the bond between the lower end cap and the media. ...causing premature failures...
...erroneous conclusions that have dogged the industry for over 25 years: That cartridge collectors can only operate at filter ratios of 2:1 or 2 fpm The assumption was that they could increase the capacity of this cartridge collector by squeezing more pleats in the cartridge...
The flaw in this reasoning is there are two factors which determine the filtering capacity of a pulse jet collector. The square foot of the media that are cleanable and the volume and pressure characteristics developed by the cleaning jet. It can only clean continuously on line when the volume is at least three times the filtering volume for each cartridge and the pressure developed by the jet is at least three times the operating pressure drop. The offline cleaning characteristics are not considered in this chapter. This applies to all pulse jet collectors regardless of the shape of the filter elements.

On the left side we see that the jet grows at an angle ... until it is stopped ... As it enters the cavity inside a filter element it forms a compression wave... The action cleans the element evenly...
...early designs the element had a truncated cone built into the closed end cap... On many dusts this would partially damage the cake... some dust would leak into the clean air side ... which led to premature filter element replacement...
...if the pulse pipe was too low, the jet would strike the media below the opening in the filter element. The cleaning action above this intersection had limited effectiveness... because of this incorrect height, the added media contributed little to the filter’s performance, but did raise the cost of constructing and operating the filters.
... once a bridge forms in the valley of the pleat, it renders the media below the bridge un-cleanable by on line cleaning.
If a cleaning system is actuated by a pressure control, the setting must be determined by experimentation since it is virtually impossible to predict the best pressure setting before the collector is installed. If it is set at a higher pressure this will affect operating pressure, frequency of cleaning (air consumption) and dust penetration through the filter. Typically a collector pressure switch is set at 3 ½ w.c. when the proper setting might be 2 “ w.c. Below are typical operating parameters at the correct and incorrect settings for a typical operation venting a material handling operation:
Operating at Correct setting for 4,000 CFM system with 3 grains per cubic foot load:
Pressure Drop 2 inch w.c
Air Consumption at 85 psig 1.5 SCFM
Dust Penetration 0.00004grains/cuft.
Cartridge life 18 months
Operating at Incorrect setting for4,000 CFM system with 3 grains per cu. ft. load:
Pressure drop 3.5 inch water column
Air consumption at 85psig 4.5 SCFM
Dust Penetration 0.00012 grains/cu.ft.
Cartridge life 6 months.
Notice that operating power is 1.32 higher, air consumption and dust emissions are increased 3 times, cartridge life is only 1/3.
Improvements in Cartridge Filter Design
Adhesives
The encapsulating compounds were replaced by adhesives. Epoxies and various
thermo-setting adhesives were supplied. This resulted in a stronger stiffer
cartridge. Cartridges were able to be constructed without the expanded outer
core. To keep the pleats from inverting during the cleaning process, plastic
bands or even strings were applied with great success. These stronger and
more open cartridges opened up new applications for cartridge collectors.
Media Changes
To go along with the changes in adhesives stronger cellulose media was developed.
They were reinforced with polyester and other fibers. This increased the
tensile strength by three or four times and allowed the introduction of
more powerful cleaning systems to increase flow per filter element. Another
important development was incorporation of pleat spacers, by upsetting the
tips of the pleats. This solved a problem in constructing cartridges by
eliminating wide variations in pleat spacing. A section of pleats that were
too narrow would bridge too easily.
Seals
In the quest to prevent any leakage from the dirty to the clean side of
the cartridge, resilient seals were added. The new seals could maintain
an even pressure on the sealing surfaces. In order to maintain a good seal
the pressure must remain within a narrow range of pressure. They can be
tightened so much that sealing compounds will go beyond their plastic limit
and produce uneven pressure and actually leak through the joint. The natural
tendency for many maintenance people is to over tighten gaskets and seals.
Some gaskets are designed to be squeezed beyond their elastic limit and
flow along the joint. This is similar to the gaskets applied in gasoline
engines. .However the valve cover gasket is really an elastic seal. In cartridge
mounting systems, the gaskets would need so much pressure that the cartridge
would be damaged. For this reason elastic seals must be applied. Stops are
often built into the seals themselves. Refer to the diagram below
L M A resilient seal, labelled L is placed between the sealing service presumably round, and at 120 degree intervals the seal is slit and a ball bearing, labelled M, is dropped into the slot. This maintains sufficient sealing pressure but does not allow over compression of the seal. Other stops can be incorporated into the mounting system.
Un-cleaned media considerations
The area of filter media that can be cleaned depends on the reverse air
volume and pressure of the reverse air jet. If we consider different orifices
and valves and assuming a permeability of 18 -20 CFM at 0.5 inches of pressure
drop for a sq. ft of the base media, the area of media cleaned is:
¼ “orifice 7-10 sq. ft.
3/8 “orifice 15-20 sq.ft.
7/16” orifice 22-26 sq. ft.
½ inch orifice 28-37 sq. ft.
¾ inch valve 75 to 78 sq.ft
1 inch valve 210 to 220 sq. ft.
If we consider a typical tandem design with two cartridges, each being cleaned by a single ¾ inch valve. Only about 80 square feet of media are cleaned. We find that the media can hold 0.1 to 0.3 lbs of mineral dust if it is not cleaned. We can compute the weight of dust on the un-cleaned media.
450 sq. ft. total less 80 sq ft. cleaned by pulse jet = 370 sq. ft.
370 sq. ft. x (0.1 to 0.3 lb.) per sq. ft. = 37 to 110 lb per 2 cartridge
tandem cleaning sets.
On some of the later designs the mounting bolts on the cartridge support tripods had to be reinforced to prevent mounting failures.
Cartridge Construction Improvements
Some other innovations to improve cartridge construction were the following.
Many new designs included a flat bottom on the closed end cap.
Some designs incorporated a flat disc above the truncated cone
Spun bond and other medias were introduced to counter the tendency of cellulose medias to expand and shrink with changes in humidity. Other cartridges were supplied with pleated medias even woven or felted media and were quite successful. With these new constructions pleated elements with wide pleat spacing operated at filter ratios of over 10.
Flexible Media pleated filter elements One limitation of some spun bond and felted medias was that they would have poor operation if the pleats were too narrow or lacked mechanical support of the media itself.
Support Media squeezed so it touches Initial final squeezed shape The support can be a laminate that is stiff but a very open media. This allows the use of virtually any media in pleated construction..
The users and designers of dust collectors have the possibility of virtually unlimited filter element life at negligible dust penetration. If the seals are intact and no dust penetration occurs to the clean side, cartridge filters can be restored to ‘as new’ condition.
|
Copyright Quality Air Management, div of Ultra-Flow Inc. © 1999 - 2008 All rights reserved. Consultants Serving Canada and the United States of America (USA) |
|
Home |
Products | Featured Articles
| Product Specifications
| Contact Us | Site Map
| Links |
Blog
|